Photo by Jason Clark (@Tacoma_EP)Photo by Jason Clark (@Tacoma_EP)Photo by Jason Clark (@Tacoma_EP)Photo by Jason Clark (@Tacoma_EP)Photo by Jason Clark (@Tacoma_EP)Photo by Jason Clark (@Tacoma_EP)Photo by Jason Clark (@Tacoma_EP)Photo by Jason Clark (@Tacoma_EP)Photo by Jason Clark (@Tacoma_EP)Photo by Jason Clark (@Tacoma_EP)
#EPeeps. Beautiful 49ms pre-QRS signal at target location. Hard to not get excited about a signal like this.
1
9
351
Redo VT. Mapped to the same location as the previous ablation. Mid-myocardial CI, OM Heatsink? Terminated during Ablation & non-inducible after ablating all moderate scar in the area. Great outcomes and teamwork on another Tandemheart VT! @GDandamudiMD #epeeps @AbbottCardio
7
30
2,746
Window settings can have meaning in macro-reentry if windows are set correctly. Signals confirmed CI & termination in 1 burn resulted. Important to remember block is the endpoint. #Abbottcardio #EPeeps #HDGrid #SensiTherm
6
21
1,847
Unexpected… Non-Ischemic VT ablation. VT LAT showed LAF/LPF BO. CS moved to visualize HBE and showed HIS CL changes before V CL changes. Entrained from LA, mapped RB potentials with #HDGrid. 1burn and non-inducible. @AbbottCardio #EPeeps #VTablation #BBRVT @GDandamudiMD
7
24
2,453
Huge news!
Breaking 🗞️ – Our #AVEIR DR Dual Chamber Leadless #Pacemaker System is now FDA approved! Discover how this first-of-its-kind #MedTech is setting the pace: bit.ly/3qQQYvx Safety Info: bit.ly/3JKOh5l
3
397
Great acute outcomes from a TandemHeart epicardial VT yesterday. High level of execution from the physicians & staff in conjunction with an impedance based mapping system ultimately led to a 3rd burn termination. @GDandamudiMD @AbbottCardio #EPeeps #VTablation (unedited geo)
1
7
1
41
4,936
(1/2) Geo creation- LV Endo is assigned to the Left Group, CS & Coronary anatomy assigned to the Right group, Epicardium assigned to the Other group (fill = 100 & not low/high as 3d lesions were used instead of 2D projection). #HDGrid used to collect LV/Epi geo.
1
1
381
CTI AFL with a good demonstration of signals at an entrance site (NF Early & FF Late), Critical Isthmus (fractionation spanning timing between signals), and exit site (FF Early & NF late about equal in timing of entrance site signals). #EPeeps
4
11
PVC this morning originating from the L/R Commissure. Good way to start the day! #EPeeps
‘Focal late’ is great piece of information. We identified focal late postero-septal in this case prompting us to make sure detailed lateral TVA mapping was performed. With focal rhythms in a healthy heart ‘focal late’ is commonly directly cross chamber to EA/BO. #EPeeps
4
16
#EPeeps. It’s important to not only use a mapping system to help with a diagnosis and end-point, but to help with patient safety as well. EnSite Split-Screen used to monitor ST segments (Pre-shadow) when burning in the LCC/RCC area.
2
5
48
A bit of a surprise. Induced by straight pacing RA @ 200ms. @AbbottCardio #EPeeps @gopi_gdanda1
2
2
9
Replying to @GDandamudiMD
@gopi_gdanda1 Ischemic VT tonight with #HDGrid. Small channel mapped. Induced VT2 while swapping for ablate, entrainment proved the VT2 target was the suspected VT1 target. No VT inducible after ablation. 2.3hr case. Great mid-week challenge.
2
10
2
44
#HDGrid @gopi_gdanda1 35min case, 32000pt HD Grid map, 281ms on AL Papillary Muscle spanning all of diastole, 0.7s of ablation while ramping up to 50W on burn #1 and no more VT. Great way to start the weekend!
11
31
4
148
Always nice to see a 1-burn termination. Always important to connect anchors with the end-point being a line of block created.
8
36
Going for ILAM & LAVA in NSR. One beautiful channel through dense endocardial scar. Confirmed as ablation target with Pacemap matching documented VT.
12
46
#EPeeps @Hapa_EP @gopi_gdanda1 Interesting use of ICE to monitor response to PNS during ablation. ViewFlex observing the liver from the IVC with Doppler. Audible changes as the right phrenic is effected with ablation.
1
11
1
25
Great PM Video!
Replying to @mattaustein
Thanks! It's all in my pacemapping video! 😉👍 youtu.be/N2hRuYifFG4
1
Over 8 VTs and could not get any complete maps without changing rhythms (unmappable??). LAT & P2P Bipolar Voltage with NSR and 3 VTs shown here. @Hapa_EP @gopi_gdanda1
3
7
1
15
Copied all of the maps to one new map for a combined fractionation & combined Voltage map. Terminated a VT on the first burn and finished a scar to scar line (4 lesions total) through the area identified with the highest fractionation. Luck or a new way to map the unmappable?
1
2
12
Interesting recent display of #HDGrid during an atypical AFL. Ablate signals at termination site vs Grid signals at termination site with equal gain, filtering, and sweep speed. CS activation was misleading and not reflecting the LA Activation Sequence along its position.
1
7